Overview

The Digital Wide Range Gauge (nWRG) family offers the capability of single port pressure measurement in the range atmosphere to 10-9 mbar, with a linear output. It’s a compact solution, halving the space and connectivity hardware requirement, which can be all important in many applications. The WRG has many novel features, including a patented striker, pushbutton calibration and set point controls and comprehensive diagnostics. The WRG is a cost-effective vacuum management solution when used either with a Edwards controller or directly integrated into the vacuum system controls.
d14790010 - nWRG RS485 NW25

nWRG RS485 NW25

D14790010
Downloading pricing...
List Price:
Web Price:
Please contact customer care for pricing
d14790020 - nWRG RS485 DN40 CF

nWRG RS485 DN40 CF

D14790020
Downloading pricing...
List Price:
Web Price:
Please contact customer care for pricing
d14790510 - nWRG RS232 NW25

nWRG RS232 NW25

D14790510
Downloading pricing...
List Price:
Web Price:
Please contact customer care for pricing
d14791010 - nWRG-L RS485 NW25

nWRG-L RS485 NW25

D14791010
Downloading pricing...
List Price:
Web Price:
Please contact customer care for pricing
d14791020 - nWRG-L RS485 DN40 CF

nWRG-L RS485 DN40 CF

D14791020
Downloading pricing...
List Price:
Web Price:
Please contact customer care for pricing
d14791510 - nWRG-L RS232 NW25

nWRG-L RS232 NW25

D14791510
Downloading pricing...
List Price:
Web Price:
Please contact customer care for pricing
d14791520 - nWRG-L RS232 DN40 CF

nWRG-L RS232 DN40 CF

D14791520
Downloading pricing...
List Price:
Web Price:
Please contact customer care for pricing

Features and Benefits

·       Digital interface removes need for Controller as can connect directly to PC or PLC

·       Magnetron uses an advanced patented technique for highly reliable striking, even at high vacuum or in relatively contaminated conditions

·       Extended service intervals and gauge lifetime, through reduction of HT voltage after ignition and low discharge current

·       Ultra compact, featuring recessed electrical connector

·       Easy service; both Pirani and magnetron elements are individually replaceable

 

Technical Data

Mass

 

NW25 version

 0.8 kg

Internal volume

26 cm3

Enclosure rating

IP40

Performance

 

Measurement range

Atmosphere to 10-9 mbar

Accuracy Typically

±15% <100 mbar

 

±30% <10-3 mbar

Maximum over-pressure

6 bar absolute

Operating and Storage Conditions

 

Temperature range

 

Operating

5 to 60 oC

Storage

0 to 70 oC

Bake-out with electronics removed

150 oC

Humidity

80% RH up to 31 oC decreasing linearly to 50% RH at 40 oC and above

Maximum altitude

3000 m

Electrical Data

 

Electrical supply voltage

15 to 48 V DC nominal

Power consumption

2 W

Set-point

open collector transistor

Rating

48 V DC 100 mA

 

Dimensions